国产精品原创剧情av-俺也去五月婷婷六月丁香-久久99热视频在线观看-国产成人精品av三区

本站熱搜:氧化鋯氧量分析儀,盤裝式氧量分析儀,電磁流量計,WSS雙金屬溫度計
  • 技術(shù)文章ARTICLE

    您當前的位置:首頁 > 技術(shù)文章 > Common faults and causes of zirconia oxygen analyzers

    Common faults and causes of zirconia oxygen analyzers

    發(fā)布時間: 2024-12-13  點擊次數(shù): 131次

    1. System maintenance.

    If the oxygen indicator shows zero or no oxygen indication in the oxygen quantity converter, first check whether the oxygen indication circuit open. If the circuit is normal, check whether the zirconia probe is normal. At a temperature of 700℃, measure the DC millolt value of the oxygen cell. If the potential is greater than 1 millivolt and the resistance is greater than 100Ω, it indicates the probe is aged and needs to be replaced with a sensor or probe. If the potential is less than 1 millivolt and the resistance is less than100Ω, it indicates an open circuit in the wiring, which should be checked and handled one by one.

    2. constant temperature system failure.

    (1) If the heating temperature cannot be maintained, check whether there is a short circuit or open circuit in the resistance wire inside the heating furnace The normal resistance of the heating furnace wire is about 60-80Ω. If the resistance wire is open or the resistance measured by the multimeter less than 60Ω, the heating furnace wire should be replaced.

    (2) When the converter shows that the temperature is too high or out of control, it not only causes inaccurate oxygen display but also acceler the aging or damage of the zirconia head. At this point, you should immediately check if the thermocouple is open-circuited, or a multimeter to check the thermocouple. If the resistance is greater than 20Ω, it is judged that the thermocouple is open-cuited, and the thermocouple and zirconia head should be replaced; another reason for the temperature失控is: relay or temperature control circuit failure, relay or temperature control circuit failure will cause the oxygen converter to always be in one of the heating voltages of the heating furnace, thus leading to an increase in.

    3. oxygen cell fault.

    (1) After the oxygen cell has been used for a period of time, the contact points of its leads may become oxidized, resulting in contact or detachment. This can cause the Em value (in units of a multimeter) to fluctuate and become unstable. Observe the oxygen cell and take action.

    (2) When the oxygen signal deviation is too large or inaccurate, it may be due to a short circuit in the zirconia electrode, which case the wiring of the zirconia cell should be checked for short circuits. When the temperature does not rise, if the internal resistance is less than 0Ω, it indicates a short circuit in the electrode, and the zirconia head needs to be replaced.

    (3) During the operation of the oxygen measurement device, if the background potential exceeds 10mV or is less than -30mV, it must be recalibrated. If the background potential remains abnormal after calibration, it indicates that the probe has aged and can no longer be used.

    4. pneumatic circuit failure.

    (1) The zirconia head is clogged with ash, causing an obstruction in the reference gas circulation. This results in a decrease in probe output potential Em and an increase in the background potential Eo. In this case, blowing several times with a bulb syringe at the gas inlet of the will resolve the issue after the gas line is unclogged.

    (2) Oxygen cell ash blockage: The probe should be removed from the ash blockage and the filter should be reinstalled.

    (3) Probe leakage has a significant impact on the probe.

    5. The oxygen content on both sides of the boiler is too high.

    By adjusting the boiler combustion conditions, when starting or stopping the boiler fan or coal mill group, if the oxygen content deviation on both sides of the boiler is large, it is a normal phenomenon. Once the boiler combustion conditions stabilize, the oxygen content display will return to normal. If the unit adjustment is not in place and the deviation is large on both sides, it is necessary to check if the zirconia head is working properly by dismantling the transmitter for inspection. The device requires the operator to adjust the gas flow. The gaseous and reference state gases should continuously flow through the electrode without stopping, and there should be no dead ends. However, too large a flow will have a cooling effect, making the cell temperature uneven and causing errors. If the flow is too small, it is not only easy to form stagnant gas flow and dead zones but also may cause local changes in oxygen concentration due to the conduction of oxygen ions, thus affecting the accuracy of the measurement and causing deviations on both sides.

    6. On-site interference causes failures.

    For newly installed instruments, due to the strong electromagnetic field, when an unstable fault occurs, the first thing to consider is the interference issue.


產(chǎn)品中心 Products
在線客服 聯(lián)系方式 二維碼

服務熱線

0550-2409223

掃一掃,關(guān)注我們

皖公網(wǎng)安備 34118102000850號

国产精品岛国片在线观看| 日本在线视频精品一区| 青青草免费在线播放视频网站| 91免费在线观看高清| 亚洲国产精品性色av| 亚洲一区二区日本乱码| 久久精品国语少妇对白| 精品国产一区二区三区四不卡在线 | 亚洲家庭伦理在线观看| 九九九热在线观看视频| 免费观看久久黄色大片| 亚洲午夜一区二区不卡| 一区二区三区四区中文在线 | 午夜精品久久久免费视频| 人妇乱系列中文字幕人妻| 亚洲91精品一区二区三区| 中文字幕乱码亚州精品一区| 日韩一区中文字幕久久| 黑人巨大一区二区三区| 欧美日韩加勒比综合在线 | 久久久亚洲熟妇熟女一区二区| 久久午夜视频在线观看| 国产精品成人免费久久黄| 亚洲国产视频中文字幕| 国产999精品在线观看| 久久精品视频就在久久| 99亚洲伊人久久精品影院| 又黄又爽区一区二区三| 视频播放一区二区三区毛片| 亚洲中文乱码一区二区| 久久精品国产亚洲熟女| 美女性生活免费视频网站| 人妻人人澡人人添人人爽桃色| 日韩电影在线播放中文字幕| 亚洲日本欧美激情综合| 亚洲丰满老熟女激情av| 国产91在线一区精品| 一区二区精品人妻av| 日韩黄av在线免费观看| 国产女主播精品视频一区| 国产91精品网站在线|